Yes, but the appropriate measurement method depends on the physical properties of the sample.
Micro-volume measurement requires a stable liquid column. Methanol and other low-surface-tension solvents may make it difficult to form a stable liquid column with a 1 µL sample. If this occurs, try increasing the sample volume to 2 µL.
If a stable liquid column still cannot be formed, use the cuvette mode on EzDrop 1000C, which does not rely on liquid column formation on the micro-volume pedestal.
Printing is handled through a computer. To print measurement results or spectra, first save them using one of the following methods:
• PC mode: Use the dedicated software to export measurement data as a CSV file. To save a spectrum image, capture the software screen, then print the saved image from the computer.
• Standalone mode: Save a screenshot of the measurement screen to a USB drive, then open and print the image on a computer.
The CSV export contains numerical data, not spectrum images.
*Screenshot availability: EzDrop 1000 units with serial number BRED-2405XXX or later can support the Screenshot function through a software update. All EzDrop 1000C units can be updated to support this function.
Printing is handled through a computer. To print measurement results or spectra, first save them using one of the following methods:
• PC mode: Use the dedicated software to export measurement data as a CSV file. To save a spectrum image, capture the software screen, then print the saved image from the computer.
• Standalone mode: Save a screenshot of the measurement screen to a USB drive, then open and print the image on a computer.
The CSV export contains numerical data, not spectrum images.
*Screenshot availability: EzDrop 1000 units with serial number BRED-2405XXX or later can support the Screenshot function through a software update. All EzDrop 1000C units can be updated to support this function.